A Chloride Ion Selective Electrode sensor specifically designed to measure 1.8 ppm to Saturated (5x10-6 M to Saturated) concentrations of Cl-.
Applications include, but not limited to: River and Tap Water, Plant Tissue, Soils, Boiler Feed Water, Blood Electrolytes/Clinical Analysis, Sweat, Urine, Cement, Plating Baths, Food Samples.
Model
Specifications
CS6210Cl CS6510Cl CS6710Cl Chloride ( Cl- ) A Chloride Ion Selective Electrode sensor
Lab Process Process Slope 56 ± 3 mV between 10-100 ppm @ 25°C
  PG13.5 NPT 3/4" Reproducibility ± 2%
CS6210Cl A Chloride Ion Selective Electrode sensor CS6510Cl A Chloride Ion Selective Electrode sensor CS6710Cl A Chloride Ion Selective Electrode sensor Interferences S-2, I-1, CN-1, Br-1
Complexation Heavy metals
Temperature Range 0 to 80° C (if calibrated at that temp.)
Pressure Range 0 to 70 psi
Response Time 95% response in 30 seconds
Storage Long term, store dry;Short term, store in dilute chloride standard
Concentration Range 1.8 ppm to Saturated
 (5 x 10-5 M to Saturated)
pH Range 2 to 12 pH
Resistance < 1 megaohm
Care, Maintenance & Cleaning Polish with polishing strips
Treat with a dilute ISA
Isopotential Point 50,000 ppm
On-Line Capability Yes
Applications River and Tap Water, Plant Tissue, Soils, Boiler Feed Water, Blood Electrolytes/Clinical Analysis, Sweat, Urine, Blood, Cement, Plating Baths, Food Samples
Site Meter