An Iodide Ion Selective Electrode sensor specifically designed to measure 0.006 ppm to Saturated (5x10-8 M to Saturated) concentrations of I-.
Applications include, but not limited to: Milk, Feeds and Plants, Pharmaceuticals.
Model
Specifications
CS6210I CS6510I CS6710I Iodide ( I- ) A Iodide Ion Selective Electrode sensor
Lab Process Process Slope 56 ± 3 mV between 10-100 ppm 25°C
  PG13.5 NPT 3/4" Reproducibility ± 2%
CS6210I A Iodide Ion Selective Electrode sensor CS6510I A Iodide Ion Selective Electrode sensor CS6710I A Iodide Ion Selective Electrode sensor Interferences S-2, Br-, CN-, Cl-, S2O3-, NH3
Complexation Heavy metals
Temp. Range 0 to 80° C (if calibrated at that temp.)
Pressure Range 0 to 70 psi
Response Time 95% response in 30 seconds
Storage Long term, store dry;Short term,
store in dilute iodide standard
Concentration Range 0.006 ppm to Saturated  
(5 x 10-8 M to Saturated)
pH Range 0 to 14 pH
Resistance < 1 megaohm
Care, Maintenance & Cleaning Polish with polishing strips
Treat with a dilute ISA
Isopotential Point 50,000 ppm
On-Line Capability Yes
Applications Milk, Feeds, Plants, Pharmaceuticals
Site Meter