A Fluoride Ion Selective Electrode sensor specifically designed to measure .02 ppm to saturated (1x10-6 M to saturated) concentrations of F-.
Applications include, but not limited to: Drinking Water, Wastewater and Natural Waters, Air and Stack Gases, Acids, Sea Water, Minerals, Soil, Foods, Biological Fluids, Toothpaste.
Model
Specifications
CS6210F CS6510F CS6710F Fluoride ( F- ) A Fluoride Ion Selective Electrode sensor
Lab Process Process Slope 56 ± 2 mV between 10-100 ppm @ 25°C
  PG13.5 NPT 3/4" Reproducibility ± 2%
CS6210F A Fluoride Ion Selective Electrode sensor CS6510F A Fluoridee Ion Selective Electrode sensor CS6710F A Fluoride Ion Selective Electrode sensor Interferences OH-1
Complexation Al+3, Fe+3, H+1
Temp. Range 0 to 80° C (if calibrated at that temp.)
Pressure Range 0 to 70 psi
Response Time 95% response in 30 seconds
Storage Long term, store dry; Short term, store in dilute fluoride standard with TISAB added
Concentration Range 0.02 ppm to Saturated  
(1 x 10-6 M to Saturated)
pH Range 5 to 8 pH
Resistance 100 to 200 kiloohms
Care, Maintenance & Cleaning Polish with polishing strips
Treat with a dilute ISA
Isopotential Point 20 ppm
On-Line Capability Yes, pH control required
Applications Drinking Water, Wastewater and Natural Waters, Air and Stack Gases, Acids, Sea Water, Minerals, Soils, Foods, Biological Fluids, Toothpaste